Netech Delta 3000A Defib / Pacer Analyzer with ECG Simulator



WARRANTY:  One Year



OVERVIEW


The Delta 3000A is a compact instrument that tests defibrillators and transcutaneous pacemakers. It is also a 12 lead ECG arrhythmia simulator.

The instrument tests defibrillators with monophasic or bi-phasic waveforms and all AEDs. It will test transcutaneous pacemkaers and includes 12 built-in user selectable test loads. The 12 lead ECG simulator includes 12 arrhythmias plus sine, square, and triangle performance waveforms. The Delta 3000A may be easily upgraded to add the capability of testing transvenous pacemakers. Operation of the Delta 3000A is via easy to follow menus and eight tactile feel keys. Instructions and measurements are displayed on the large four line LCD display. A print key outputs test measurements to the RS 232 serial port.

The Delta 3000A is a complete instrument. It is delivered in a hard carrying case with paddle adapters and two set of universal connectors, an AC adapter, batteries, an instruction manual, a Certificate of Calibration, and the Netech two year warranty. The Delta 3000A is a unique instrument that meets the demanding requirements of the biomedical engineer in a rugged, light weight package. Its upgradeability places it in a class clearly beyond the competition.

 

HIGHLIGHTS


          Compact, Lightweight Case

          Simple, Menu Driven Operation

          Large Easy to read LCD Display

          Simultaneous Display of All Test Measurements

          Dedicated Print Key for Immediate Documentaion

          RS 232 Interface

          Battery or AC Operation

          One-Year Warranty


MEASUREMENTS


Defibrillator Tester:- Tests all Defibrillators - Semi, Automatic, and those with Bi-Phasic         Waveforms.


         Measures: Energy, Peak Voltage, Peak Current, and Cardioversion.

         Storage and Playback of output waveforms

         Built-in 50 Ohm Test Load.

         Built-in User Selectable Test Loads to 1000 Ohms.


Pacemaker Analyzer:- Tests All Transthoracic Pacemakers.


         Built-in 50 Ohm Test Load

         Built-in User Selectable Test Loads to 1000 Ohms.

         Amplitude, Rate, Pulse Width Accuracy Tests.

         Sensed and Paced Refactory Period Tests.

         50 and 60 Hz Immunity Tests.

         Measures Pulse Energy.


ECG Simulator:- Tests all ECG Equipment.


         12 lead ECG.

         Performance Waveforms: Sine, Square, Triangle.

         12 Arrhythmias.

         8 Selectable Rates and Amplitudes


STANDARD ACCESSORIES


         Hard Carrying Case

         AC Adapter

         Universal Paddle Adapter

         Shrouded Banana Plugs (set of 2)

         User Manual

         Certificate of Calibration

         Instruction Manual


OPTIONAL ACCESSORIES


         Interface cables for various Maufacturers

         Palm Interface Software




SPECIFICATIONS:


Defibrillator Analyzer


Energy Measurement: Load: 50 Ohm 1 %

Range (High): 900 J, Res.1 J

Resolution: 1 Joules

Range (Low): 0 - 100

Resolution: 0.1 Joules

Accuracy: +/- 2% of the reading

+ 1 J for Hi Range or

+ 0.5 J for Low Range

Maximum Voltage: 5200 V

Maximum Current: 100 Amp (H)

Sync. Time: 0 - 250 ms

Waveform Output:

Real Time: Output 1000 : 1 Stored.

Discharge waveform expanded 200:1 and is repeated 8 sec. It can be viewed through the paddles and through the ECG High Level Output


Pacemaker Analyzer


Transcutaneous:

Test Load: 50 Ohm (Fixed) and variable 100 to 1000 Ohms user selectable

Current: 1 to 200 mA

Accuracy: +/- 5% + 0.5 mA

Pulse Width: 0.5 to 80 ms

Accuracy: +/- 1% + 0.1 ms

Refractory: 20 - 500 ms

Accuracy: +/- 2% or + 10 ms

Atrial and Ventricular:

Test Load: 500 Ohms

Current: 1 to 25 mA

Accuracy: +/- 5% + 0.5 mA

Pulse Rate: 30 to 800 ppm.

Accuracy: +/- 1% + 0.1 ms

Refactory: +/- 20 - 500 ms

Accuracy: +/- 2% or 10 ms

Immunity Test: 50 / 60 Hz


ECG ArrhythmiaSimulator


ECG: Normal Sinus Rhythm.

Rate: 30, 60, 90, 120, 240

Performance: Sine, Square, Triangle and Pulse

Rate: 0.5, 1, 2, 10, 15, 20, 25, 40 Hz

Arrhythmias:

VFBC, VFBF, VTAC, AFIB, PVC, PVC1, PVC2, BGY, RBB, AFIB, ATRFT, NSR

Waveform Output: 12 Lead ECG and Paddle and High Level Output Jacks.



General


Power Requirements:

Two 9V Alkaline Batteries Or AC Adapter


Dimension:

10 x 11 x 2.25 inches

(25.4 x 27.9 x 5.7 cm)

Weight: 4 lbs (1.8 kg)